Deformation of internal thread of dental implants on repeated abutment screw tightening

  • Mohammed Novman Post graduate student, Department of Prosthodontics and Implantology, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, INDIA
  • Divya Hegde Professor, Department of Prosthodontics and Implantology, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, INDIA
  • Lakshmikanth K Associate Professor, Department of Prosthodontics and Implantology, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, INDIA
  • Chandrasekharan K Nair Professor and Head of the Department of Prosthodontics and Implantology, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, INDIA
  • Sadhvi K V Senior Lecturer, Department of Prosthodontics and Implantology, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, INDIA

Abstract

Objective: To evaluate the surface changes of internal threads of the implant that occurs after repeated tightening and loosening of the abutment screw.
Methodology: MIS implant analogue of dimension 13x3.5 mm were mounted. Light body impression was made to copy the internal threads of dental implant before abutment screw tightening. The impression surface was then studied under a stereomicroscope to visualize the surface of implant threads. Abutment screw was tightened and removed eight times from the implant, followed by making a light body impression of the internal threaded surface of the implant to visually analyse the surface changes under stereomicroscope.
Results: The distance between four threads measured before and after screw tightening using Progres image pro software, the values were before screw tightening 1.923 mm and 1.861 and after screw tightening 1.901 mm 1.843 respectively. No significant difference was observed.
Conclusions: It can be concluded that there is no change in the internal threads of the implant on repeated screw tightening and loosening.

Published
2013-06-29
How to Cite
NOVMAN, Mohammed et al. Deformation of internal thread of dental implants on repeated abutment screw tightening. Trends in Prosthodontics and Dental Implantology, [S.l.], v. 4, n. 1, p. 20-21, june 2013. ISSN 0975-9298. Available at: <https://trendsinprosthodontics.com/index.php/trends/article/view/55>. Date accessed: 24 sep. 2021.
Section
Articles

Most read articles by the same author(s)

1 2 > >>